Back to Test Catalogue

Cystatin C with Estimated Glomerular Filtration Rate (eGFR), Serum/Plasma

Test ID: CYSTC

Test

Method

Immunoturbidimetric

Platform

Roche Cobas c501

Report Includes

Cystatin C; Cystatin C eGFR

Specimens

Serum; Plasma (Li heparin)

Test Location

Children's Hospital Eastern Ontario, 401 Smyth Road, Ottawa ON K1H 8L1

Test Version

26-May-2023

Specimen

Specimens

Serum; Plasma (Li heparin)

Sample Volume

1 mL

Collection & Handling

Handling Information

Separate and freeze immediately. Store and send frozen.

Stability

Refrigerated 7 days
Frozen See below

Additional Stability Information

Stability at -20°C or lower: Plasma 6 months or serum 24 months.
If the specimen thaws, it is unsuitable for analysis.

Rejection Criteria

Specimen Not frozen

Test Version

26-May-2023

Performance / Interpretation

Method

Immunoturbidimetric

Platform

Roche Cobas c501

Turnaround Time

5 days

Results

Name Units Reference Range Conversion Factor
Cystatin C mg/L
  • ≤ 1.20
mg/dL x 10
CYSCeGFR mL/min/1.73m2
  • ≥ 90
    CYSCeGFR = GlomerularFiltration Rate estimatedfrom measured Cystatin C)
    NOTE: Accuracy of eGFR formula has not been validated for eGFR ≥90 mL/min/1.73m2
For anyone <18 years of age, eGFR will be calculated using the Schwartz Cystatin C Equation (Kidney Int. 2012 Aug;82(4):445-53).
For anyone ≥18 years of age, eGFR will be calculated using the CKD-EPI Cystatin C Equation (NEJM. 2012 Jul5;367(1):20-9).

Test Location

Children's Hospital Eastern Ontario, 401 Smyth Road, Ottawa ON K1H 8L1

Test Version

26-May-2023

Interface / Setup

HL7 Interface Codes

Order Code Result Codes Units
CYSTC 28596Cystatin C mg/mL
28597CYSTCeGFR mL/min/1.73m2

Test Version

26-May-2023