Test Overview
Method
Inductively coupled plasma-triple quadrupole mass spectrometry (ICP-TripleQ MS)
Result Included
Titanium
Specimen
Specimen Type
Avoid gel separator tubesPlasma (K2-EDTA)
Containers
Preferred Containers
BD 6 ml K2-EDTA Royal Blue Vacutainer (Reference # 368381)
Volume
Sample Volume
2 mL
Specimen Comment
Avoid gel separator tubes.
Collection & Handling
Collection Instructions
Collect in contaminant-free tube.
Handling Information
Separate as soon as possible and transfer plasma to polypropylene vial. Store and send cold.
Additional Information
Concentrations of titanium are higher in erythrocytes than in plasma. The results in plasma may be falsely elevated if not separated within 30 m and/or hemolysis is present.
Rejection Criteria
| Criteria | Specification |
|---|---|
| Hemolysis | Visible |
Performance & Interpretation
Turnaround Time
10 days
Results
Name
Units
Reference Range
Conversion
-
Titaniumnmol/L0.0 - 7.9ng/L x 0.0209
Referral Location
Canada
Test Version
Last Updated