Procollagen Type I Intact N-Terminal Propeptide, Serum

Test ID: PINP

Test

Aliases/Synonyms

PINP I

Method

RIA (Radioimmunoassay)

Method Description

A known amount of labelled PINP and an unknown amount of unlabelled PINP in the sample compete for a limited number of high affinity binding sites on the polyclonal rabbit anti-PINP antibody. A second antibody, directed against rabbit IgG and coated to kaolin particles, is used to separate the antibody-bound PINP from free PINP. The radioactivity of the bound tracer antigen is measured in a gamma counter. The amount of labelled PINP in the sample tube is inversely proportional to the amount of PINP in the sample. The concentrations in unknown samples are obtained from a calibration curve, which is based on the concurrent testing of the UniQ PINP RIA calibrators that contain 0 to 250 µg/L PINP.

Platform

Orion Diagnostica UniQ PINP

Report Includes

Procollagen I Intact N-Terminal

Specimens

Serum

Clinical Utility

Intact N-terminal propeptide of type I procollagen is useful in the management of osteoporosis and monitoring bone formation therapies and antiresorptive therapies.

Test Location

Quest Diagnostics, California USA

Test Version

21-Sep-2023

Specimen

Specimens

Serum

Sample Volume

1.0 mL

Minimum Volume

0.3 mL

Stability

Ambient Unacceptable
Refrigerated 27 days
Frozen 60 days

Additional Stability Information

Avoid repeated freezing and thawing.

Test Version

21-Sep-2023

Performance / Interpretation

Method

RIA (Radioimmunoassay)

Method Description

A known amount of labelled PINP and an unknown amount of unlabelled PINP in the sample compete for a limited number of high affinity binding sites on the polyclonal rabbit anti-PINP antibody. A second antibody, directed against rabbit IgG and coated to kaolin particles, is used to separate the antibody-bound PINP from free PINP. The radioactivity of the bound tracer antigen is measured in a gamma counter. The amount of labelled PINP in the sample tube is inversely proportional to the amount of PINP in the sample. The concentrations in unknown samples are obtained from a calibration curve, which is based on the concurrent testing of the UniQ PINP RIA calibrators that contain 0 to 250 µg/L PINP.

Platform

Orion Diagnostica UniQ PINP

Turnaround Time

8 days

Results

Name Units Reference Range Conversion Factor
Procollagen I Intact N-Terminal mcg/L
  • Males:
  • <23 y: Not established
  • 23 - 60 y: 30-110
  • ≥60 y: Not established
  • Females:
  • <20 y: Not established
  • 20 - 45 y: 22 - 104
  • 46-60y: 20-108
  • ≥60 y: Not established

Test Location

Quest Diagnostics, California USA

Test Version

21-Sep-2023

Interface / Setup

Test Version

21-Sep-2023