Share to
  • Email
  • Facebook
  • X
  • Bluesky
  • Instagram
  • LinkedIn
Share to
  • Email
  • Facebook
  • X
  • Bluesky
  • Instagram
  • LinkedIn
Back to Test Catalogue
Test ID: TITAN P

Titanium, Plasma

Test Overview

Method

High resolution sector field inductively coupled plasma-mass spectrometry (HR-SF-ICP-MS)

Result Included

Titanium (µg/L); Titanium (nmol/L)

Specimen

Specimen Type

Avoid gel separator tubes
Plasma (K2-EDTA)

Containers

Preferred Containers

BD 6 ml K2-EDTA Royal Blue Vacutainer (Reference # 368381)

Volume

Sample Volume

3.0 mL

Minimum Volume

1.0 mL

Specimen Comment

Avoid gel separator tubes.

Collection & Handling

Collection Instructions

Collect in contaminant-free tube. Avoid hemolysis.

Handling Information

Separate as soon as possible and transfer plasma to polypropylene vial. Store and send cold.

Additional Information

Concentrations of titanium are higher in erythrocytes than in plasma. The results may be falsely elevated if not separated within 30 m and/or hemolysis is present.

Stability

Ambient Refrigerated Frozen
22 days 14 months 14 months

Rejection Criteria

Criteria Specification
Hemolysis Visible

Performance & Interpretation

Turnaround Time

10 days

Results

Name
Units
Reference Range
Conversion
  1. Titanium
    µg/L
    < 3.00
  2. Titanium
    nmol/L
    < 62.7

Test Version

Last Updated

2024-07-19